Objective
Our purpose is to review results of AMS 800 artificial urinary sphincter implantation for treatment of male stress urinary incontinence.
To evaluate indications, contraindications, patient assessment, operative technique and complications of AMS 800 device.
Materials and methods
We used up to date literature of PubMed database.
Results
Success rates of artificial urinary sphincter AMS 800 vary between 61% and 100%. 80 – 90% of patients were satisfied with the
outcome and would recommend having the surgery. The complication rates are generally low and varies between 1–31%.
AMS 800 5-year functioning without complications rates vary between 59% and 79%.
Conclusions
AMS 800 artificial urinary sphincter is highly effective and time tested treatment of moderate and severe stress urinary incontinence.